Light scattering from a diffractive-refractive intraocular lens: a goniometer-based approach for individual zone assessment

Grzegorz Łabuz, Thomas J T P van den Berg, Gerd U Auffarth, Ramin Khoramnia

Onderzoeksoutput: Bijdrage aan wetenschappelijk tijdschrift/periodieke uitgaveArtikelWetenschappelijkpeer review

9 Citaten (Scopus)
63 Downloads (Pure)

Samenvatting

We proposed and tested a method to measure light scattering from the diffractive lens profile in an echelle element featuring 9 zones. Measurements were performed using a goniometer-based setup up to 7.5°. The proportion of scattered light was calculated to derive the loss of light. Material scattering was minimal (∼1 deg2/sr); however, each echelle zone acted as a scattering source. A nearly gradual straylight increase was found with the zone number showing peak intensity between 3° and 3.75°. An estimated 6.2% ± 0.1% was lost due to scattering, which ought to be considered when reporting an IOL's light loss.

Originele taal-2Engels
Pagina's (van-tot)6724-6732
Aantal pagina's9
TijdschriftBiomedical Optics Express
Volume13